Restore Barbieri Spectro LFP
Problem/Question
"LFP is an OEM device and can´t be controlled"
The
Barbieri Spectro LFP is not recognized in Ergosoft and the Barbieri
Gateway displays an error message stating that the Barbieri Spectro
LFP is an OEM device and can´t be controlled.
Solution/Work-Around
The device itself has lost the saved parameters and must be restored using the procedure described below.
- Serial of the device (located on the Barbieri Spectro LFP itself, e.g. B311974).
- Service Report of the device (to create it please follow these instructions)
- Open Barbieri Gateway Software (available for free download from the Barbieri Service Support Page)
- Select "Tools"
- Click on "Get Service Report"
- Save the created file
After
having sent the information described above, the Barbieri Support will
send you a generated txt file with the respective parameters and an
associated procedure to restore the device.
After this procedure should the Spectro LFP work as usual.
Related Articles
0x1203 Error: Command failed: S1 when measuring with Barbieri Spectro LFP
Problem/Question When attempting to measure and/or do the White calibration with the Barbieri Spectro LFP, the measuring head does not move and it gives out the error message "0x1203 Error: Command failed: S1". Solution/Work-Around This issue usually ...
Barbieri Spectrophotometer, Epson printer not available in the driver list
Problem/Question When setting up a Barbieri Spectrophotometer or an Epson printer, the model is missing in the device list. The HTM drivers for Epson printers as well as the Barbieri Gateway software are large installation files. In order to keep the ...
Missing Barbieri Measurement Tool Selection
Problem/Question After an update to the current version 16, the Barbieri measurment tool is missing in the measuring tool selection. Since Barbieri has extremely enlarged its Gateway installer, we have decided to make it available as an additional ...
Error 1200 The specified device name is invalid
Problem/Question After a job is screened and sent to the device, this error message occurs: "Error 1200 The specified device name is invalid" This problem usually occurs when something has been changed or renamed in an existing PrintQueue. ...
Error when sending data to "Port name"