Restore Barbieri Spectro LFP

Restore Barbieri Spectro LFP

Problem/Question
"LFP is an OEM device and can´t be controlled"
The Barbieri Spectro LFP is not recognized in Ergosoft and the Barbieri Gateway displays an error message stating that the Barbieri Spectro LFP is an OEM device and can´t be controlled.

Solution/Work-Around
The device itself has lost the saved parameters and must be restored using the procedure described below.
Please contact the support of Barbieri (support@barbierielectronic.com) providing them with the following information:
  1. Serial of the device (located on the Barbieri Spectro LFP itself, e.g. B311974).
  2. Service Report of the device (to create it please follow these instructions)
  1. Open Barbieri Gateway Software (available for free download from the Barbieri Service Support Page)
  2. Select "Tools"
  3. Click on "Get Service Report"
  4. Save the created file
After having sent the information described above, the Barbieri Support will send you a generated txt file with the respective parameters and an associated procedure to restore the device.
After this procedure should the Spectro LFP work as usual.

    • Related Articles

    • 0x1203 Error: Command failed: S1 when measuring with Barbieri Spectro LFP

      Problem/Question When attempting to measure and/or do the White calibration with the Barbieri Spectro LFP, the measuring head does not move and it gives out the error message "0x1203 Error: Command failed: S1". Solution/Work-Around This issue usually ...
    • Barbieri Spectrophotometer, Epson printer not available in the driver list

      Problem/Question When setting up a Barbieri Spectrophotometer or an Epson printer, the model is missing in the device list. The HTM drivers for Epson printers as well as the Barbieri Gateway software are large installation files. In order to keep the ...
    • Missing Barbieri Measurement Tool Selection

      Problem/Question After an update to the current version 16, the Barbieri measurment tool is missing in the measuring tool selection. Since Barbieri has extremely enlarged its Gateway installer, we have decided to make it available as an additional ...
    • Error 1200 The specified device name is invalid

      Problem/Question After a job is screened and sent to the device, this error message occurs: "Error 1200 The specified device name is invalid"   This problem usually occurs when something has been changed or renamed in an existing PrintQueue. ...
    • Error when sending data to "Port name"

      Problem/Question When sending print data, the following error message appears from the Print/CutQueue “Error when sending data to "Port name". The process hosting the driver for this device has ended." This error message indicates that there are ...